Transmittion of Heart Sounds by Means of Auscultation
DOI:
https://doi.org/10.20535/RADAP.2020.80.73-78Keywords:
auscultation of heart sounds, second heart tone, electronic stethoscope/phonendoscope, conductive tube, mechanical impedance of human body part, electronic stethoscope, electronic phonendoscopeAbstract
The article presents the results of the study of the amplitude-frequency characteristics of the electron stethoscope / phonendoscope with a sound-conducting tube in the frequency range (50 - 400) Hz, which according to different estimates belongs to the spectrum of II tones of the human heart. Modeling is carried out taking into account mechanical impedance of the human chest area. In order to generalize the results, in contrast to the zone of registration of II tone, the area located in the fourth intercostal margin along the right middle clavicle line with known parameters was chosen. Mechanical impedance of the site, according to the measurement data, was interpreted by the system of sequentially connected elements with concentrated parameters: elasticity, friction resistance and vibrational mass. The friction resistance was determined by the minimum of the impedance modulus, the elasticity of the latter at low frequencies (50 Hz), and the vibrational mass at high frequencies (1000 Hz). The stethoscope/phonendoscope were modeled by elements with concentrated and distributed parameters. The stiffness of the phonendoscope membrane was determined by experimental measurements and the resonance frequency by calculations. The sound conducting tube was simulated by the system with distributed parameters. It was accepted that the mechanical impedance of the load of the sound-conducting tube in the frequency range under study had an elastic character, taking into account the frequency range of the microphone. At low frequencies, the input impedance of the phonendoscope prevailed over that of the stethoscope. For the rigidity of the microphone membrane of more than 103 N/m in the studied frequency range there are bands with a significant predominance of impedance of the human body or means of auscultation of heart sounds. In the first case, the heart's sound system operates in the preset pressure mode, while in the second case, it operates at an oscillatory speed. In the case of a preset pressure, the transmission functions of the stethoscope and the phonendoscope are different, while the preset oscillatory speed is similar. In the frequency bands in which the ausculators functioned at a given oscillatory speed, there was a decrease in sounds and a slight increase in the given pressure.Downloads
Published
2020-03-30
Issue
Section
Radioelectronics Medical Technologies
License
Copyright (c) 2020 Y. V. Storchun, Y. I. Yakovenko, T. A. Smerdova
This work is licensed under a Creative Commons Attribution 4.0 International License.
Authors who publish with this journal agree to the following terms:- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).